BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Getinge Washer Disinfector 8666 and 8668, Getinge Group

Product group General equipment for medical treatment
Reference 01552/11

Getinge Group has issued a field safety notice concerning a problem with not isolated overheating protector of Getinge Washer Disinfector 8666 and 8668 equipped with the water analyses option. There is an obvious possibility of inadvertently touching the non-protected electrical parts while taking a water sample which can lead to a servere electric shock, potentially causing serious injury.

Safety Notice for Getinge Washer Disinfector 8666 and 8668, Getinge Group , Download_VeroeffentlichtAm_EN PDF, 60KB, File does not meet accessibility standards