Field safety notice for slings (produced for ProLift S222 sitto-stand hoist), Novacare

Product group General equipment for medical treatment - ward equipment and equipment for emergency medical services
Reference 05604/13

The company Novacare GmbH has issued a field safety notice concerning the slings produced for the ProLift S222 sitto-stand hoist as from September 2012. Some slings have not been correctly sewn. This could Iead to the patient falling backwards out of the hoist and therefore a high security risk exists.

Field safety notice for slings (produced for ProLift S222 sitto-stand hoist), Novacare , Download_VeroeffentlichtAm_EN PDF, 845KB, File does not meet accessibility standards