Lot Recall for Pinnacle Cure Sleeves, KERR
2011.11.11
Product group Dental products
Reference 04892/11
Pinnacle Cure Sleeves: The company KERR has informed about a lot recall of these dental light guide sleeves. The action was initiated because of a packaging error in such a way that sleeves for 7 or 8 mm light guides were labeled as sleeves for 11 mm light guides. If used, the devices would not achive complete coverage of the light guide.
Lot Recall for Pinnacle Cure Sleeves, KERR , Download_VeroeffentlichtAm_EN