BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Medical Device Recall: NC / RC Screws for CARES Abutment, STRAUMANN

Product group Dental products - dental implants
Reference 04591/12

NC / RC Screws for CARES Abutment: The company STRAUMANN has informed about a lot recall of these dental implant abutment components. The action was initiated because of a labelling error in such a way that devices with specification ‘NC Screw’ have been marked as ‘RC Screw’ and vice versa.

Medical Device Recall: NC / RC Screws for CARES Abutment, STRAUMANN , Download_VeroeffentlichtAm_EN PDF, 44KB, File does not meet accessibility standards