BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall of the Vario Knee System Resection Guide Femoral Distal from Plus Orthopedics

Product group Medical instruments for use in humans
Reference 03151/05

Recall of the VKS (Vario Knee System) Resection Guide Femoral Distal (article number 1214122, Index A, lots PI.79 and PI.80) by Plus Orthopedics AG. There is a risk, that the weld holding the superior adjustment knob fails, whereby the arm separates from the distal femur resection guide.

Advisory Notice is only available in German language.

Recall of the Vario Knee System Resection Guide Femoral Distal from Plus Orthopedics , Download_VeroeffentlichtAm_EN PDF, 23KB,