Recall for NexGen Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor, Zimmer Inc.

Product group Medical instruments for use in humans
Reference 02470/08

Recall of the products NexGen Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (catalogue numbers and lots see annex). The instruments are prone to fracture at the tip during use when subjected to non-axial impaction loads. There is the risk, that the metal fragment from the fracture is being left in the patient.

Recall for NexGen Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor, Zimmer Inc. , Download_VeroeffentlichtAm_EN PDF, 59KB, File does not meet accessibility standards