Safety Notice for the RT-Plus Modular surgical technique, Smith & Nephew
2012.05.11
Product group Medical instruments for use in humans - bone surgery
Reference 03210/12
Smith & Nephew revised the RT-Plus Modular surgical technique instructions. Stepped Drill and Drill Guide should only used with manual driven handle. Powertool use may cause that the instruments get jammed and cause metal debris or bone defects.