Venous Line Clamp Occlusion Failure on hemodialysis instruments SYSTEM 1000, TINA, AURORA and ARENA from Baxter

Product group Injections / Infusions / Transfusions / Dialysis
Reference 03373/05

The company Baxter informed in an Important Product Information of a possible Venous Line Clamp Occlusion Failure on hemodialysis instruments SYSTEM 1000, TINA, AURORA and ARENA. Baxter discovered that fluid or air can pass through the venous line what could result in an air emboli condition. Users are requested to check the pinch zone of the venous line clamp for a correct tubing position.