Information for the Fiamm Sealed Lead Acid Batteries in IVAC® P Series Infusion Pumps from Cardinal Health Alaris® Products

Product group Injections / Infusions / Transfusions / Dialysis
Reference 01085/06

The company Cardinal Health Alaris® Products notified customers of Fiamm Sealed Lead Acid Batteries in IVAC® P Series Infusion Pumps of a possible premature failure. Customers have been informed to replace Fiamm batteries with an alternative type when transporting patients longer than 30 minutes.

Information for the Fiamm Sealed Lead Acid Batteries in IVAC® P Series Infusion Pumps from Cardinal Health Alaris® Products , Download_VeroeffentlichtAm_EN PDF, 21KB, File does not meet accessibility standards