BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information about B|Braun Space Perfuso from B. Braun Melsungen

Product group Injections / Infusions / Transfusions / Dialysis
Reference 02686/06

B. Braun Melsungen has issued a customers information concerning the B|Braun Space Perfusor- and Infusomat with prime function. After the prime process is finished the occlusion pressure does not automatically fall back to the manual pressure setting but remains at the maximum.

Information about B|Braun Space Perfuso from B. Braun Melsungen , Download_VeroeffentlichtAm_EN PDF, 56KB, File does not meet accessibility standards