Recall of the Arctic Circler® Linear Cardiac Cryoablation Catheter from CryoCath Technologies

Product group Injections / Infusions / Transfusions / Dialysis
Reference 03008/06

The company CryoCath Technologies Inc. issued an information concerning the recall of the Arctic Circler® Linear Cardiac Cryoablation Catheter. Due to a twist in the catheter shaft a catheter pressurization in conjunction with gas embolism is possible.

Recall of the Arctic Circler® Linear Cardiac Cryoablation Catheter from CryoCath Technologies , Download_VeroeffentlichtAm_EN PDF, 97KB, File does not meet accessibility standards