BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall of the Level 1 Normothermic I. V. Fluid Administration Set from Smiths Medical

Product group Injections / Infusions / Transfusions / Dialysis
Reference 03396/06

Smiths Medical ASD Inc. recalls several Lots of the Level 1 Normothermic I. V. Fluid Administration Set from the market. Due to a blockage of the recirculating warming fluid chanel non-tempered infusion fluid may be administered. The fluid warmer alarms and shutsdown automatically.

Recall of the Level 1 Normothermic I. V. Fluid Administration Set from Smiths Medical , Download_VeroeffentlichtAm_EN PDF, 103KB, File does not meet accessibility standards