BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information concerning the Level 1 from Smiths Medical

Product group Injections / Infusions / Transfusions / Dialysis
Reference 01001/07

Smiths Medical has issued an information that when the Level 1 is switched off or the power is otherwise lost the clamp of the air detector is deactivated. This could allow any air within the patient line to be delivered to the patient.

Information concerning the Level 1 from Smiths Medical , Download_VeroeffentlichtAm_EN PDF, 105KB, File does not meet accessibility standards