BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information for Alaris® GP, Cardinal Health

Product group Injections / Infusions / Transfusions / Dialysis
Reference 04536/07

The company Cardinal Health issued a Field Safety Notification on the Alaris® GP volumetric pump concerning a software upgrade. In a rare sequence of events either the single bubble and/or accumulator air detection system can be temporarily disabled. Until the upgrade process is undertaken the manufacturer recommends some measures to minimize the risk.

Information for Alaris® GP, Cardinal Health , Download_VeroeffentlichtAm_EN PDF, 54KB, File does not meet accessibility standards