Safety Information about Blood Leak Detectors of single 4008 & 5008 dialysis system, Fresenius Medical Care

Product group Injections / Infusions / Transfusions / Dialysis
Reference 01312/08

The company Fresenius Medical Care performs a replacement of Blood Leak Detectors in 4008 and 5008 dialysis systems. In some instances predamaged blood leak detector circuit boards have been assembled leading to an undetected blood loss into the dialysate circuit.

Safety Information about Blood Leak Detectors of single 4008 & 5008 dialysis system, Fresenius Medical Care , Download_VeroeffentlichtAm_EN PDF, 46KB, File does not meet accessibility standards