BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall Ultra filters U 8000 S, Gambro

Product group Injections / Infusions / Transfusions / Dialysis
Reference 01900/08

The Gambro Dialysatoren GmbH recalls one lot of the Ultra filters U 8000 S from the market. Due to the cleaning process with hot / cold water cycles the filter may become leaky. The leakage may result in a misbalancing of the patient during dialysis.

Recall Ultra filters U 8000 S, Gambro , Download_VeroeffentlichtAm_EN PDF, 43KB, File does not meet accessibility standards