Recall BD 60 ml Luer-Lok Syringes, Medtronic

Product group Injections / Infusions / Transfusions / Dialysis
Reference 03125/08

Recall of the BD 60 ml Luer-Lok Syringes in Magellan Platelet Separator Disposable Products and Stand Alone Syringes (Models see annex, all lots). The unit package seal integrity and resulting product sterility of the syringes can be adversely affected when the product is exposed to low atmospheric pressure.

Recall BD 60 ml Luer-Lok Syringes, Medtronic , Download_VeroeffentlichtAm_EN PDF, 51KB, File does not meet accessibility standards