BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Alaris® PK Syringe Pump, Cardinal Health

Product group Injections / Infusions / Transfusions / Dialysis
Reference 00094/09

The company Cardinal Health has issued a Field Safety Notice concerning a display anomaly affecting Alaris® PK Syringe Pumps with software versions v3.2.15 and earlier. Under certain conditions an erroneous induction dose is displayed on the INDUCTION CONFIRM screen in TCI mode. A software upgrade v3.2.16 will be available.

Safety Notice for Alaris® PK Syringe Pump, Cardinal Health , Download_VeroeffentlichtAm_EN PDF, 142KB, File does not meet accessibility standards