Field Safety Notice for dialysis systems Gambro AK 200 S equipped with UFD and AK 200 ULTRA S in combination with BiCart Select, Gambro

Product group Injections / Infusions / Transfusions / Dialysis
Reference 01956/09

The company Gambro Lundia AB issued a Field Safety Notice for dialysis systems Gambro AK 200 S equipped with UFD and AK 200 ULTRA S in combination with BiCart Select. The machines may freeze during functional check. If the machine was left in this state for more than one hour, an unwanted high concentration of potassium might occur.

Field Safety Notice for dialysis systems Gambro AK 200 S equipped with UFD and AK 200 ULTRA S in combination with BiCart Select, Gambro , Download_VeroeffentlichtAm_EN PDF, 66KB, File does not meet accessibility standards