BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for LifeCare 4200 PCA Infusion System, Hospira Inc.

Product group Injections / Infusions / Transfusions / Dialysis - infusion technology
Reference 05570/12

Hospira has issued a Field Safety Notice concerning manipulations during use of PCA Infusion Systems. Failure to close the slide clamp whilst manipulating the syringe (vial) or stopping the infusion, may result in an over delivery of the medication being infused.

Safety Notice for LifeCare 4200 PCA Infusion System, Hospira Inc. , Download_VeroeffentlichtAm_EN PDF, 444KB, File does not meet accessibility standards