Recall of certain lots of Plastipak Syringes, Becton Dickinson

Product group Injections / Infusions / Transfusions / Dialysis - syringes and cannulas
Reference 01973/13

The manufacturer Becton Dickinson informs about the recall of certain lots of Plastipak Syringes, product references 300223, 300865 und 300869, with old design stopper (larger stopper). If syringes with old design stopper are used in syringe pumps that have been adjusted for the new stopper, there may be a delay in initiating an occlusion alarm.

Recall of certain lots of Plastipak Syringes, Becton Dickinson , Download_VeroeffentlichtAm_EN PDF, 90KB, File does not meet accessibility standards