Safety Notice for Alaris® GP Infusion System, CareFusion Switzerland
2013.01.08
Product group Injections / Infusions / Transfusions / Dialysis - infusion technology
Reference 07078/12
CareFusion has issued a Field Safety Notice concerning the problem that the Alaris® GP Infusion System produces very small air bubbles (less than 50µl) which are too small to trigger the lowest configurable air-in-line detection thresholds (alarm) of the Alaris GP Infusion System
Safety Notice for Alaris® GP Infusion System, CareFusion Switzerland , Download_VeroeffentlichtAm_EN