Corrective action for the enteral feeding pump Amika, Fresenius Kabi
2014.01.06
Product group Injections / Infusions / Transfusions / Dialysis - infusion technology
Reference 07418/13
The company Fresenius Kabi notified of a corrective action for the enteral feeding pump Amika. It is possible that pumps equipped with software version 2.0.d and 2.0.e will not trigger the air in line / empty bag alarm. A new software version 2.1.c has been developed.
Corrective action for the enteral feeding pump Amika, Fresenius Kabi , Download_VeroeffentlichtAm_EN