Corrective action of Becton Dickinson for FACSCanto Flow cytometer

Product group In-vitro diagnostics
Reference 02499/04

Corrective action of Becton Dickinson for FACSCanto Flow cytometer with FACSDiva Software Version 4.0.1.2 or lower. The manufacturer informed the customers that a software defect can result in a drop in sheath fluid pressure if acquisition is started without a properly loaded tube. Customers were advized to follow exactly the instructions of the instrument's user guide in order toprevent the problem A corrected softwareversion will be provided at a later time.