BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Corrective action of Ortho-Clinical Diagnostics for Vitros CRP Microslides

Product group In-vitro diagnostics
Reference 02934/04

Corrective action of Ortho-Clinical Diagnostics for Vitros CRP Microslides (Coating batches 401-409). The affected batches show a reduced linearity range which may cause a negative bias at high analyte concentrations. The customers were informed about the problem and how to avoid it.

Corrective action of Ortho-Clinical Diagnostics for Vitros CRP Microslides , Download_VeroeffentlichtAm_EN PDF, 119KB,