Corrective action of Cell-Dyn 3700 CS / Cell-Dyn 3700 SL, Cell-Dyn 3500 CS / Cell-Dyn 3500 SL, Cell-Dyn 3000 CS / Cell-Dyn 3000 SL from Abbott
2007.09.25
Product group In-vitro diagnostics
Reference 01855/07
Corrective action of Abbott for Cell-Dyn 3700 CS / Cell-Dyn 3700 SL, Cell-Dyn 3500 CS / Cell-Dyn 3500 SL, Cell-Dyn 3000 CS / Cell-Dyn 3000 SL (all serial numbers) because of potentially failure of hemoglobin measurement due to electrostatic discharge. Analyzers will be upgraded per mandatory Technical Service Bulletin.