Recall for PATHWAY anti-HER2/neu (4B5) Rabbit Monoclonal Primary Antibody, Roche
2009.09.25
Product group In-vitro diagnostics
Reference 03295/09
Recall of Roche Diagnostics GmbH (Ventana Medical Systems) for PATHWAY anti-HER2/neu (4B5) Rabbit Monoclonal Primary Antibody (PATHWAY HER2 (4B5)), lot no. 680957 because of light staining. The customers were informed about the problem and requested to stop using the lot affected or to perform retesting of samples run with the affected lot.