Corrective action for ADVIA Chemistry Systems Ethanol_2 Reagent, Siemens

Product group In-vitro diagnostics
Reference 04765/10

Corrective action of Siemens for ADVIA Chemistry Systems Ethanol_2 Reagent because of revision to on-board calibration and reagent stability claims because the Reagent Blank (RBL) will reach the current high limit at an earlier time than stated in the method Instructions for Use. The customers were informed about the problem and how to avoid it.

Corrective action for ADVIA Chemistry Systems Ethanol_2 Reagent, Siemens , Download_VeroeffentlichtAm_EN PDF, 24KB, File does not meet accessibility standards