BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Leonardo Workstation, Siemens

Product group Medical data processing (software)
Reference 00511/08

Siemens AG has issued a Safety Notice concerning Leonardo Workstation with InSpace 3D and software VB30B. If volume data are reconstructed based on an acquisition series that was acquired with a patient orientation other than “Head first – Supine”, the resulting slice images show wrong orientation labels. A software update is planned.

Safety Notice for Leonardo Workstation, Siemens , Download_VeroeffentlichtAm_EN PDF, 68KB, File does not meet accessibility standards