Medtronic is replacing older version QUIK-COMBO pacing/defibrillation/ECG cables and Standard paddles

Product group Medical electronics / Electromedical devices
Reference 02236/05

Medtronic ERS is replacing older version QUIK-COMBO pacing/defibrillation/ECG (therapy) cables and/or Standard (hard) paddles used with the Lifepak 12 defibrillator/monitor series.The connector pins could break during patient use. Medtronic recommends to leave the Lifepak 12 in use until these accessories are replaced as well as performing regular routine testing.

Medtronic is replacing older version QUIK-COMBO pacing/defibrillation/ECG cables and Standard paddles , Download_VeroeffentlichtAm_EN PDF, 25KB,