BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall of the Respiratory Monitors Nico Model 7300 from Respironics

Product group Medical electronics / Electromedical devices
Reference 02592/06

Respironics Critical Care has issued information concerning a recall of the Respiratory Monitors Nico Model 7300 with the serial numbers beginning with 138. There is a risk that the audio alert speaker component fails resulting in no audible alarm during an alert condition. The affected monitors are re-equipped with a Speaker Upgrade Kit and a Software Update 5.3, 21-08-06.

Recall of the Respiratory Monitors Nico Model 7300 from Respironics , Download_VeroeffentlichtAm_EN