Recall of HeartStart FR2+ Automated External Defibrillators, models M3860A, M3861A, M3840A and M3841A from Philips Medizin Systeme Böblingen

Product group Medical electronics / Electromedical devices
Reference 00225/07

The Philips Medizin Systeme Böblingen GmbH has issued an information concerning a recall of HeartStart FR2+ Automated External Defibrillators, models M3860A, M3861A, M3840A and M3841A, manufactured between late 2004 and early 2005. Button(s), the On/Off, Shock and/or option buttons, on the front panel may not respond to an initial button press and may require additional button presses or additional force to activate the button.