Information concerning Lifeline semiautomatic external defibrillator, model DDU-100 from Defibtech
2007.03.20
Product group Medical electronics / Electromedical devices
Reference 00733/07
Defibtech, LLC has issued an information concerning a corrective action of the Lifeline semiautomatic external defibrillator, model DDU-100 Series. There is the risk that the self-test software may incorrectly clear a low battery warning. The manufacturer advises the affected customers to perform a test procedure immediately and then once per month until a modified software has been installed.