Recall LeVeen SuperSlim Needle Electrodes, Boston Scientific

Product group Medical electronics / Electromedical devices
Reference 01748/08

The company Boston Scientific Corporation is conducting a Medical Device Recall of the LeVeen SuperSlim Needle Electrodes. The cannula of the electrode may become detached from its correct orientation inside the handle and may result in the tines not being fully retracted. Clinical effects are possible.

Recall LeVeen SuperSlim Needle Electrodes, Boston Scientific , Download_VeroeffentlichtAm_EN PDF, 66KB, File does not meet accessibility standards