Corrective action for Primedic Charger Basis, Metrax
2008.06.02
Product group Medical electronics / Electromedical devices
Reference 01958/08
The Metrax GmbH has issued an information concerning a corrective action of the Primedic Charger Basis for external defibrillators. By using the device without attached defibrillator it may lead to an inflated leakage current at the charging connections. By mounting of two protective caps a direct contact to the charging connections is prevented.
Corrective action for Primedic Charger Basis, Metrax , Download_VeroeffentlichtAm_EN