Information for Lifepak CR Plus defibrillator / Lifepak Express defibrillator, Medtronic
2008.08.05
Product group Medical electronics / Electromedical devices
Reference 02870/08
The Medtronic Physio-Control has issued an information concerning a corrective action of the Lifepak CR Plus and Lifepak Express automated external defibrillators. A potential short circuit may result in preventing the defibrillator from powering on although the OK indicator was visible on the Readiness Display. Affected customers are recommended immediately to inspect the defibrillators to ensure their devices are functional.