BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Field Safety Notice concerning a Corpuls 3 Defibrillator/Monitoring systems

Product group Medical electronics / Electromedical devices
Reference 03838/08

GS Elektromedizinische Geräte G. Stemple GmbH has issued a Field Safety Notice concerning a Corpuls 3 Defibrillator/Monitoring systems series software issue up to and including software version 1.4.2. Under certain conditions the Monitoring Unit may indicate the message “Pairing failed” with the consequence that the device is not ready for use. The new software verison 1.4.3 corrects the error.

Field Safety Notice concerning a Corpuls 3 Defibrillator/Monitoring systems , Download_VeroeffentlichtAm_EN PDF, 221KB, File does not meet accessibility standards