BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Corrective action for Responder 2000 defibrillator/monitor, Cardiac Science Corp.

Product group Medical electronics / Electromedical devices
Reference 04003/08

The Cardiac Science Corp. has issued a Field Safety Notice concerning a corrective action of defibrillator/monitor Responder 2000. Pressing and holding down the shock button for more than 5 seconds causes a “Button Stuck” error which can result in therapy prohibition. An Operator’s Manual addendum has been released containing details and instructions on how to prevent this error.

Corrective action for Responder 2000 defibrillator/monitor, Cardiac Science Corp. , Download_VeroeffentlichtAm_EN PDF, 77KB, File does not meet accessibility standards