BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for PowerHeart AED G3, Cardiac Science Corp.

Product group Medical electronics / Electromedical devices
Reference 00166/09

Cardiac Science Corp. has issued a Field Safety Notice concerning a corrective action of the external defibrillators PowerHeart AED G3. Due to an error of the German G3 Language Pack the affected devices will issue a voice prompt “Schock abgeben / deliver shock” after delivering a shock instead of “Schock abgegeben / shock delivered.
The customers will receive a software update which will correct this the users confusing voice prompt as soon as available. Advisory Notice is only available in German language.

Safety Notice for PowerHeart AED G3, Cardiac Science Corp. , Download_VeroeffentlichtAm_EN PDF, 90KB, File does not meet accessibility standards