Corrective action for optional NIBP units, installed in defibrillators cardioLife, Nihon Kohden

Product group Medical electronics / Electromedical devices
Reference 04709/09

Nihon Kohden Europe GmbH has issued a Field Safety Notice concerning a corrective action of optional NIBP units models CY-0010 and SG-761VK, installed in defibrillators cardioLife TEC-7621K, TEC-7631K, TEC-7721K and TEC-7731K for monitoring of noninvasive blood pressure. Due to a malfunction by setting the alarm limits in the neonatal operation mode the alarm is impaired or is prevented completely.