Safety Notice for TruSat Pulse Oximeter with external power supply, GE Healthcare Finland Oy
2010.04.29
Product group Medical electronics / Electromedical devices
Reference 01455/10
GE Healthcare has issued a Field Safety Notice that the TruSat Pulse Oximeter with external power supply does not comply with current safety standards. It is possible that the external power supply could leak electrical current. This current could cause third degree burns or an abnormal heart rhythm that, if untreated, could lead to death.