BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Field Safety Notice concerning BeneHeart D6 defibrillator / monitor, Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Product group Medical electronics / Electromedical devices
Reference 03470/10

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. has issued a Field Safety Notice concerning a corrective action of the BeneHeart D6 defibrillator / monitor. A high voltage (HV) printed circuit board assembly (PCBA) may contain a component with excess solder which might create a short. There is the risk that affected devices do not deliver therapy. In Germany, the product is not in the market.

Field Safety Notice concerning BeneHeart D6 defibrillator / monitor, Shenzhen Mindray Bio-Medical Electronics Co., Ltd. , Download_VeroeffentlichtAm_EN PDF, 1MB, File does not meet accessibility standards