Safety Notice for Transport Pro Monitor, GE Healthcare
2010.12.01
Product group Medical electronics / Electromedical devices
Reference 04397/10
GE Healthcare informs about a safety issue that can occur when using the Transport Pro Monitor with the CARESCAPE Patient Data Module (PDM).
Transport Pro stops communication with PDM after 414 days of continuous run time. This results in the loss of the displayed parameter information (e.g. waveforms, numerical data, alarms).
This condition affects all released versions of software.