Safety Notice for PRIMEDIC™ AkuPak, Metrax GmbH
2010.12.28
Product group Medical electronics / Electromedical devices
Reference 04399/10
Metrax GmbH has issued a Field Safety Notice concerning a corrective action of the storage battery PRIMEDIC™ AkuPak 73765 which are designed for use with external defibrillators of the product series Primedic™ M250 and Primedic™ M290. Due to an error in the software an incorrect battery charge status may result in the equipment switching over to an unexpected “power-off” state during an emergency treatment. For implementation of the software update the affected storage batteries have to be returned to Metrax.
Safety Notice for PRIMEDIC™ AkuPak, Metrax GmbH , Download_VeroeffentlichtAm_EN