BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for PRIMEDIC™ AkuPak, Metrax GmbH

Product group Medical electronics / Electromedical devices
Reference 04399/10

Metrax GmbH has issued a Field Safety Notice concerning a corrective action of the storage battery PRIMEDIC™ AkuPak 73765 which are designed for use with external defibrillators of the product series Primedic™ M250 and Primedic™ M290. Due to an error in the software an incorrect battery charge status may result in the equipment switching over to an unexpected “power-off” state during an emergency treatment. For implementation of the software update the affected storage batteries have to be returned to Metrax.

Safety Notice for PRIMEDIC™ AkuPak, Metrax GmbH , Download_VeroeffentlichtAm_EN PDF, 52KB, File does not meet accessibility standards