Safety Notice: Corrective Action for Meducore Easy, Weinmann Geräte für Medizin GmbH + Co. KG
2011.04.18
Product group Medical electronics / Electromedical devices
Reference 00323/11
Weinmann Geräte für Medizin GmbH + Co. KG has issued a Field Safety Notice concerning a corrective action of the external defibrillator Meducore Easy. During the safety check it can occur that the time allotted for charging the capacitor is insufficient to charge it completely. The defibrillator interprets this condition as a device failure and is no longer ready for use. A modified firmware will solve this problem.