Safety Notice for the Lucas 2 Chest Compression System, Medtronic Physio-Control

Product group Medical electronics / Electromedical devices
Reference 00880/11

Medtronic Physio-Control has issued a Field Safety Notice concerning a corrective action of the Lucas 2 Chest Compression Systems of the manufacturer Jolife AB produced and shipped in April 2009 through December 2010. Due to a failure of battery contact connections the devices may stop in fail safe mode and the light-emitting diodes may start flickering or the devices may switch off completely. The affected systems will be retrofitted.

Safety Notice for the Lucas 2 Chest Compression System, Medtronic Physio-Control , Download_VeroeffentlichtAm_EN PDF, 24KB, File does not meet accessibility standards