BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Patient Monitor, Model 91387, Spacelabs Healthcare Inc.

Product group Medical electronics / Electromedical devices
Reference 01758/11

Spacelabs Medical has issued a Field Safety Notice about a potential hazard associated with the use of Spacelabs Medical Patient Monitor, Model 91387, shipped from February 10, 2010 till March 24, 2011 (serial number 1387-1xxxxx or higher). The circuit that supplies backup power to the modules is not working. If there is a power interruption, the three-minute backup may not be operational and you may lose any changes you have made to the alarm settings.

Safety Notice for Patient Monitor, Model 91387, Spacelabs Healthcare Inc. , Download_VeroeffentlichtAm_EN PDF, 35KB, File does not meet accessibility standards