Safety Notice concerning a corrective action of the Zoll E Series external defibrillators, Zoll Medical

Product group Medical electronics / Electromedical devices
Reference 02011/11

Zoll Medical has issued a Field Safety Notice concerning a corrective action of the Zoll E Series external defibrillators. In case of a shock advisory in the BLS mode devices with software versions 7.10 or 7.11 may fail to auto charge the capacitor. The failure will be eliminated by a software upgrade.

Safety Notice concerning a corrective action of the Zoll E Series external defibrillators, Zoll Medical , Download_VeroeffentlichtAm_EN PDF, 58KB, File does not meet accessibility standards