BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall for the Aquamantis 2.3 Bipolar Sealer, Salient Surgical Technologies

Product group Medical electronics / Electromedical devices
Reference 03407/11

The company Salient Surgical Technologies, Inc. notified of recall for the Aquamantis 2.3 Bipolar Sealer, Model #23-113-1. The manufacturer identified a manufacturing defect in the insulation of the electrical conductor within the shaft. A burn may result, if an unintentionally exposed conductor is in contact with tissue.

Recall for the Aquamantis 2.3 Bipolar Sealer, Salient Surgical Technologies , Download_VeroeffentlichtAm_EN PDF, 33KB, File does not meet accessibility standards